Chapter 3 Software Navigation
Page 58 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Informational Icons
In addition to the icons in the Quick Navigation Bar, there are two icons that provide treatment-
related information. Refer to the following table for a description of the informational icons.
3.2 Administrative Screens
Home Screen
There are two different Home Screens: one for the
CATALYS®
System administrator and another
for general users. In this manual, the Home Screen for the
CATALYS®
System administrator is
referred to as the “Administrator Home Screen” and the Home Screen for general users is
referred to simply as the “Home Screen”.
The
CATALYS®
System administrator has access to several features that are not accessible to
other users. When the administrator logs into the system using the administrator user name and
password, the Administrator Home Screen displays. From the Administrator Home Screen, the
administrator can:
• Go to the Users Screen, All Treatments & Reports Screen, Surgeon Templates Overview
Screen, and Settings Screen
• Verify system alignment
• Add new treatment plans and edit previously added treatment plans
• Initiate a treatment plan
Icon
Description
The orientation icon represents the rotation of the selected eye relative to the
CATALYS®
System. The icon rotates to display axis markings for your reference.
The colored ring provides a visual indication of the forces exerted on the
disposable lens. The ring is green when the forces exerted on the disposable
lens are within an acceptable range to lock the suction ring/disposable lens
assembly to the system. The ring is yellow when the forces exerted on the
disposable lens rise to a level of concern. The ring is orange when the forces
exerted on the disposable lens rise to a level that will stop an OCT scan or a laser
treatment. The ring is red when the forces exerted on the disposable lens rise to
an unacceptable level.
NOTES
If the ring is orange or red during
INTEGRAL GUIDANCE
System imaging, you
must press the RESCAN EYE button to proceed.
If the ring is red during the docking process, the system will undock the patient
after 3 seconds. If the ring is red during laser treatment, the system will undock
the patient after 1 second.