Chapter 3 Software Navigation
Part Number: 0160-6411, Rev A, 02/2019
Page 129 of 314
Software Release: cOS 5.00
Verify Fluid Panel
After the system verifies that the lock is secure, the Verify Fluid panel opens. As instructed in the
panel, check the video image of the patient’s eye to ensure that the suction ring is completely
filled with sterile buffered saline solution and that no air bubbles are present. The video image
should appear sharp and clear.
Figure 3.83 Docking Screen with Verify Fluid Panel Open
After verifying that the video image of the patient’s eye is sharp and clear, press the FLUID
CONFIRMED button to initiate
INTEGRAL GUIDANCE
System imaging. A check mark will appear
next to the Verify Fluid panel after pressing the FLUID CONFIRMED. The force indicators will also
change from only showing a red band to showing yellow, orange, and red bands, indicating
different severity levels of forces being exerted by the patient on the disposable lens.
When the disposable lens, vacuum, capture, and lock steps have all been completed, the lateral
and vertical force indicators (indicator to the left of the video and on top of the video) will
minimize to the upper left corner of the video for the remainder of the treatment.
WARNING
Continuously monitor the video image immediately before and throughout each laser treatment.
Continuously verify that the suction ring remains completely filled with sterile buffered saline
solution. If any air bubbles and/or a meniscus appear on the video image before treatment, do
not initiate laser treatment. If air bubbles and/or a meniscus appear during treatment, then
terminate the laser treatment by immediately releasing the laser footswitch.