Chapter 2 System Overview
Page 44 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
5. When the system detects that the disposable lens has been properly installed, the
Vacuum panel opens. As instructed in the Vacuum panel, place and center the suction
ring of the
LIQUID OPTICS™
Interface on the plastic test hemisphere, apply patient
vacuum by depressing the right (patient vacuum ON) footswitch or by pressing the
(patient vacuum)
I
(on) button on the docking keypad, then add sterile buffered saline
solution to the suction ring.
Figure 2.15 Vacuum Panel
6. When the system detects that the suction ring has been placed and the patient vacuum
applied, the Capture panel opens. As instructed in the Capture panel, raise the suction
ring until the indicator arrow on the right is within the Capture Zone, as shown in the
following figure.
Figure 2.16 Capture Panel
7. When the indicator arrow is within the Capture Zone, press the
(Capture)
I
(on)
button on the docking keypad to capture the suction ring in the disposable lens.