Chapter 4 CATALYS® System Procedure
Part Number: 0160-6411, Rev A, 02/2019
Page 191 of 314
Software Release: cOS 5.00
2. When all three arrows in the video image are in the green zones, press the
(Lock)
I
(on) button on the docking keypad (located on the system front panel directly above the
patient) to stabilize the suction ring/disposable lens assembly with the laser.
Figure 4.35 Turn On Lock
QUICK TIP: LOCKING THE SUCTION RING
Alternatively, you can activate Guided Docking by pressing and holding the
(
Lock)
I
(on) button. The system will adjust the bed/chair until all three indicators are within their
respective Lock Zones. Release the button after the lock engages or to stop the bed/chair
at any point.
WARNING
Verify that the suction ring is correctly connected to the disposable lens component of
the
LIQUID OPTICS™
Interface during the initial patient docking procedure.
3. To confirm that the suction ring and disposable lens are secured with respect to the
laser, gently grab the handle of the suction ring and attempt to move it slightly. If the
suction ring does not move, proceed to verify fluid. If the suction ring does move, wiggle
it until it establishes a secure vacuum connection and locks into place, and then
reconfirm eye stability by gently grabbing and moving the handle of the suction ring
again.
QUICK TIP: LOCKING THE SUCTION RING
Always verify on the video image that the suction ring has snapped into place and that
any and all movements have ceased.
Lock ON
Button