Chapter 3 Software Navigation
Part Number: 0160-6411, Rev A, 02/2019
Page 131 of 314
Software Release: cOS 5.00
3.6 Treatment Screens
The treatment screens allow you to perform 3-D
INTEGRAL GUIDANCE
System imaging; to verify
and customize surface fits and treatment parameters prior to laser treatment; and to initiate
and monitor laser treatment. There are five main types of treatment screens:
• Surface Mapping Review Screen—allows you to verify and customize surface fits
• Incision Review Screens—allow you to review and verify treatment parameters
• Incision Adjustment Screens—allow you to adjust treatment parameters
• Final Review Screen—allows you to review and approve treatment parameters prior to
laser treatment
• Treatment Progress Screen—allows you to monitor laser treatment
NOTE
Refer to
on page
for a detailed description of the parameter
selections available on the treatment screens.
Surface Mapping Review Screen
After capturing and locking the suction ring; verifying that the video image of the eye is sharp
and clear; and pressing the FLUID CONFIRMED button on the final Docking Screen, the Surface
Mapping Review Screen displays, and
INTEGRAL GUIDANCE
System imaging begins
automatically. The initial Surface Mapping Review Screen displays axial and sagittal cross-section
images of the corneal surfaces. To display images of all surfaces, press the NEXT button or the
icon.
WARNING
Safety margins for all incisions are preserved only if custom fit adjustments to ocular surface(s)
are applied in accordance with the instructions for use. Purposeful misuse of custom fit
adjustments to ocular surface(s) can result in patient injury and complication(s) and, therefore,
must be avoided.
From the Surface Mapping Review Screen, you can use the automatically detected ocular
surfaces or custom fit the identified surfaces of the:
• Cornea Anterior/Posterior
• Lens Anterior/Posterior
• Pupil/Iris
• Limbus