Appendix A – Chapter 1 Mobile Patient Bed System Description
Page 244 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Refer to
on page
for detailed information on caster and
directional lock modes.
Lock Foot Pedals
The lock foot pedals are located on the left and right sides of the Mobile Patient Bed base. The
lock foot pedals activate caster lock mode when pressed. Refer to
for detailed information on caster lock mode.
Emergency Stop Button
The emergency stop button is located on the patient left side of the Mobile Patient Bed base.
The emergency stop button disables electrical power to all motors in the bed when pressed.
Refer to
Disabling the Mobile Patient Bed
for additional information on the
emergency stop feature.
NOTE
The emergency stop button is a latching switch that will remain depressed until the button is
rotated to disengage it.
Reset Button
The reset button is located next to the lock foot pedal on the patient right side of the Mobile
Patient Bed base. If the Mobile Patient Bed encounters a critical error, use a pin or paper clip to
press the reset button. Refer to
Resetting the Mobile Patient Bed
on page
for additional
information on the reset feature.
Battery
The Mobile Patient Bed is powered by 2 rechargeable 12V sealed lead-acid (SLA) batteries. The
on-board battery is designed to operate for a minimum of 2 days (at 12 treatments per day) and
can be recharged by plugging the bed's AC power cord into a standard AC wall outlet. When
recharged daily, the battery requires 2 to 4 hours to recharge. A completely discharged battery
requires 12 hours to recharge. For optimal performance, it is recommended that the battery is
charged on a nightly basis. The batteries are expected to last 2-3 years when charged as
recommended.
Batteries are not replaceable by the customer. Replacement batteries will be installed by
Technical Services. However, the batteries must be ordered and obtained by the customer prior
to installation. Batteries can be ordered from local battery distributors. Technical Services can
help identify a local distributor if needed. Used batteries must be disposed of in accordance with
local regulations. Approved batteries are:
• Panasonic LC-X1220P
• Power-Sonic PDC-12200
The
(battery status) icon on the control pendant indicates the battery status, as follows:
• Solid green—battery power is high; sufficient power for at least two treatments
• Slowly blinking green—battery power is low; sufficient power for one treatment