Appendix D – Software Error Code Table
Part Number: 0160-6411, Rev A, 02/2019
Page 307 of 314
Software Release: cOS 5.00
Host NoTreat
07-025
Automatic iris fit error, verify surface fits, consider re-scan
Host NoTreat
07-026
Capsulotomy Z out of range, re-scan and verify surface fits
Host NoTreat
07-027
Lens incision centering error, re-scan and verify surface fits
Host NoTreat
07-028
Lens Fragmentation parameter error, contact OptiMedica Service
Host NoTreat
07-029
Lens Fragmentation parameter error, contact OptiMedica Service
Host NoTreat
07-030
Lens Fragmentation parameter error, contact OptiMedica Service
Host NoTreat
07-031
Lens fragmentation Z out of range, re-scan and verify surface fits
Host NoTreat
07-032
Lens Fragmentation parameter error, contact OptiMedica Service
Host NoTreat
07-033
Lens too thin for lens fragmentation, verify surface fits, consider re-
scan
Host NoTreat
07-034
Lens fragmentation pattern too large, verify surface fits and adjust
parameters
Host NoTreat
07-035
Ramp pattern generation error, contact OptiMedica Service
Host NoTreat
07-036
Lens fragmentation pattern too large, verify surface fits and adjust
parameters
Host NoTreat
07-037
Iris intersects limbus, verify surface fits, consider re-scan
Host NoTreat
07-038
Iris intersects suction ring, verify surface fits, consider re-scan
Host NoTreat
07-039
Disposable lens scan error, verify disposable lens is correctly installed
and re-scan
Host NoTreat
07-040
Disposable lens scan error, verify disposable lens is correctly installed
and re-scan
Host NoTreat
07-041
Disposable lens scan error, verify disposable lens is correctly installed
and re-scan
Host NoTreat
07-042
Suction ring position scan error, verify suction ring is correctly docked
and re-scan
Host NoTreat
07-043
Suction ring position check error, verify suction ring is correctly
docked and re-scan
Host NoTreat
07-044
Corneal incision center rotation error, re-scan and verify surface fits
Host NoTreat
07-045
Suction ring detection error, verify suction ring fluid and position,
then re-scan
Host NoTreat
07-046
Cornea anterior scan error, re-scan and verify surface fits
Error
Type
Error
Number
Error
Message