Chapter 3 Software Navigation
Page 106 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
─ Custom—allows you to drag the touchscreen image to the desired location within the
safety zone when in the treatment screens. Dragging is only available during
treatment.
Once you have chosen valid settings, the selected arcuate incisions pattern displays in the center
of the eye model, as shown in the preceding figure. Press the DETAILS button to proceed to the
Arcuate Incisions Details Screen, the BACK button to return to the (Basic) Screen for the previous
procedure in the sequence, or the NEXT button to proceed to the (Basic) Screen for the next
procedure in the sequence.
NOTES
If you selected only the arcuate incisions procedure on the Plan a Treatment Screen, pressing the
BACK button on the Arcuate Incisions (Basic) Screen will take you back to the Plan a Treatment
Screen and pressing the NEXT button will take you directly to the Treatment Plan Summary
Screen.
If you press the HOME button on the Arcuate Incisions (Basic) Screen, the current treatment
parameters will be saved.
Arcuate Incisions Details Screen
To access the Arcuate Incisions Details Screen, press the DETAILS button on the Arcuate Incisions
(Basic) Screen or the Quick Navigation Bar Arcuate Incisions Icon on the Capsulotomy, Lens
Fragmentation, or Cataract Incisions Primary or Sideport(s) Details Screen.
Figure 3.63 Arcuate Incisions Details Screen
From the Arcuate Incisions Details Screen, you can select the following parameters:
• Penetration Type
─ Anterior Penetrating—has uncut region in posterior cornea