Chapter 2 System Overview
Part Number: 0160-6411, Rev A, 02/2019
Page 35 of 314
Software Release: cOS 5.00
(± 3.6 °F) during a 45-minute period, the system will shut down. If the temperature of the
installation environment goes below 15 °C or above 32 °C at any time, the system will shut
down. The system should be positioned such that the treatment beam is aimed away from any
windows or doors. A laser safety sign must be posted at the entrance to the treatment room.
Do not block cooling airflow or cooling vents on the laser system. Allow at least 5 cm (2 in) of
clear space around the laser system to provide adequate system cooling airflow. Use care when
routing system cables to prevent a tripping hazard. If cords or cables must cross a floor where
there is traffic, use of a floor cord/cable cover is recommended.
WARNINGS
The detachable AC power cord is the main disconnect of the equipment from mains AC power.
The detachable AC power cord must be disconnected from the system to ensure that the system
is completely isolated from mains AC power. Make sure adequate space is available during
system installation to allow access to the AC power cord, the system circuit breakers, the remote
interlock connector, and other rear panel connections.
Do not simultaneously touch accessible contacts on any conductors and the patient.
Do not replace the AC power cord, laser footswitch cable, and vacuum footswitch cable. The
CATALYS®
System is designed for use with the AC power cord, laser footswitch cable, and
vacuum footswitch cable supplied with the system. If a replacement AC power cord, laser
footswitch cable, and vacuum footswitch cable is required, contact
OPTIMEDICA
Service to
obtain the correct replacement part for your system.
Connecting the System Components
Before starting up the system, connect the
CATALYS®
System components as follows:
1. Insert the corresponding end of the power cable into the
CATALYS®
System main power
receptacle.
2. Insert the door interlock connector, laser footswitch cable, and vacuum footswitch cable
into the respective receptacles. These are locking and keyed connectors. Align the red
dot on the connector with the red dot on the receptacle, and insert the connector until it
snaps in place.