Chapter 3 Software Navigation
Page 126 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Figure 3.80 Docking Screen with Capture Panel Open—Suction Ring Outside Capture Zone
The patient is centered, but the suction ring of the
LIQUID OPTICS™
Interface is not in contact
with the disposable lens. The green bar represents the Capture Zone, and the arrow indicates the
patient’s position with respect to the Capture Zone. In this example, the bed/chair needs to be
lowered until the suction ring is in the Capture Zone and the disposable lens is engaged.
When the indicator arrow on the right side of the video is within the Capture Zone, press the
(Capture)
I
(on) button on the docking keypad to capture the suction ring in the
disposable lens. When the system detects that the suction ring has been captured, the system
emits an audible “ding” and a check mark displays in the Capture panel.
NOTE
Refer to
on page
for detailed instructions on positioning and
capturing the suction ring, as well as adjusting the bed/chair.