Appendix A – Chapter 3 Mobile Patient Bed Maintenance
Part Number: 0160-6411, Rev A, 02/2019
Page 265 of 314
Software Release: cOS 5.00
Table 3.2 Guidance and Manufacturer's Declaration: Immunity
(This table applies to IEC60601-1-2, 3rd Edition compliant Mobile Patient Bed manufactured prior to January 1st, 2019)
The Mobile Patient Bed is intended for use in the electromagnetic environment specified below. The customer or
the user of the Mobile Patient Bed should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment: Guidance
ESD
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV Contact
±15 kV Air
Floors should be wood, concrete, or
ceramic tile. If floors synthetic, the r/h
should be at least 30%.
EFT
IEC 61000-4-4
±2 kV Mains
±1 kV I/Os
±2 kV Mains
±1 kV I/Os
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Surge
IEC 61000-4-5
±1 kV Differential
±2 kV Common
±1 kV Differential
±2 kV Common
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Voltage Dips/Drop-
out IEC 61000-4-11
>95% Dip for 0.5 Cycle
>95% Dip for 1 Cycle
30% Dip for 25/30
Cycles
>95% Dip for 250/300
Cycles
>95% Dip for 0.5 Cycle
>95% Dip for 1 Cycle
30% Dip for 25/30
Cycles
>95% Dip for 250/300
Cycles
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the Mobile Patient
Bed requires continued operation during
power mains interruptions, it is recom-
mended that the Mobile Patient Bed be
powered from an uninterruptible power
supply or a battery.
Power Frequency
50/60 Hz
Magnetic Field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should
be that of a typical commercial or hospital
environment.