Agilent X-Series Signal Generators SCPI Command Reference
55
Basic Function Commands
List/Sweep Subsystem ([:SOURce])
Range
See also
[:LEVel][:IMMediate][:AMPLitude]
command for output power ranges.
Remarks
The setting enabled by this command is not affected by signal generator power–on,
preset, or
*RST
.
:LIST:POWer:POINts
Supported
All Models
[:SOURce]:LIST:POWer:POINts?
This command queries the number of power points in the current list sweep file.
:LIST:RETRace
Supported
All Models
[:SOURce]:LIST:RETRace ON|OFF|1|0
[:SOURce]:LIST:RETRace?
This command configures the sweep to retrace to the first sweep point, or stop at the last sweep
point upon completion of each sweep.
ON (1)
The sweep retraces to the first sweep point.
OFF (0)
The sweep stays at the last sweep point of the completed sweep and stays there
until sweep is initiated and triggered again. When sweep is initiated and triggered
again, the sweep point moves to the first point of the sweep.
*RST
1
Key Entry
Sweep Retrace Off On
:LIST:TRIGger:EXTernal:SOURce
Supported
All Models
:LIST:TRIGger:EXTernal:SOURce TRIGger[1]|TRIGger2|PULSe
:LIST:TRIGger:EXTernal:SOURce?
This command selects the external trigger source. With external triggering, the selected bi- directional
BNC is configured as an input.
TRIGger1
This choice selects the TRIG 1 BNC as the external trigger source for triggering
sweep, point and function generator sweeps.
TRIGger2
This choice selects the TRIG 2 BNC as the external trigger source for triggering
sweep, point and function generator sweeps.
PULSe
This choice selects the PULSE BNC as the external trigger source for triggering
sweep, point and function generator sweeps.
Example
:LIST:TRIG:EXT:SOUR PULS
The preceding example sets the external trigger source to the PULSE BNC.