402
Agilent X-Series Signal Generators SCPI Command Reference
Real–Time Commands
Custom Subsystem–Option 431([:SOURce]:RADio:CUSTom)
:STANdard:SELect
Supported
N5172B/82B with Option 431
[:SOURce]:RADio:CUSTom:STANdard:SELect NONE|AC4Fm|ACQPsk|BLUEtooth|CDPD
[:SOURce]:RADio:CUSTom:STANdard:SELect?
This command selects a predefined setup for Custom (with the appropriate defaults) and/or clears
the selection.
NONE
This choice clears the current predefined Custom format.
AC4Fm
This choice sets up an Association of Public Safety Communications Officials (APCO)
compliant, compatible 4-level frequency modulation (C4FM) format.
ACQPsk
This choice sets up an Association of Public Safety Communications Officials (APCO)
compliant, compatible quadrature phase shift keying (CQPSK) format.
BLUEtooth
This choice sets up a Bluetooth (2-level frequency shift keying) format.
CDPD
This choice sets up a minimum shift keying Cellular Digital Packet Data (CDPD)
format.
*RST
NONE
:TRIGger:TYPE
Supported
N5172B/82B with Option 431
[:SOURce]:RADio:CUSTom:TRIGger:TYPE CONTinuous|SINGle|GATE
[:SOURce]:RADio:CUSTom:TRIGger:TYPE?
This command sets the trigger type.
CONTinuous
The framed data sequence repeats continuously; the sequence restarts every time the
previous playback is completed. To customize continuous triggering, refer to
“:TRIGger:TYPE:CONTinuous[:TYPE]” on page 402
.
SINGle
The framed data sequence plays once for every trigger received.
GATE
An external trigger signal interrupts the playback while the gating signal is in the
inactive state. Playback resumes when the external control signal returns to the active
state. The active state can be set to high or low.
*RST
CONT
:TRIGger:TYPE:CONTinuous[:TYPE]
Supported
N5172B/82B with Option 431
[:SOURce]:RADio:CUSTom:TRIGger:TYPE:CONTinuous[:TYPE] FREE|TRIGger|RESet
[:SOURce]:RADio:CUSTom:TRIGger:TYPE:CONTinuous[:TYPE]?
Key Entry
None
APCO 25w/C4FM
APCO 25 w/CQPSK
Bluetooth
CDPD
Key Entry
Continuous
Single
Gated