Agilent X-Series Signal Generators SCPI Command Reference
Analog Modulation Commands
Pulse Modulation Subsystem–Option UNU and UNW and Option 320 ([:SOURce])
The optional variable <unit> accepts nS (nanoseconds) to S (seconds) with a resolution of 10 nS.
The range value is dependent on the pulse period (refer to the
:PULM:INTernal:PERiod
command).
Use DELay1 with the DOUBlet parameter and Delay1 and Delay2 with the ADOublet parameter (refer
to the
:PULM:SOURce:INTernal
command).
When “TRIGgered” is the pulse train trigger mode, then this value specifies the time after a trigger is
received before the first Pulse Train On Time starts (refer to the
:PULM:INTernal:TRAin:TRIGger
command).
Example
:PULM:INT:DEL 200E-9
The preceding example sets the internal pulse delay to 200 nanoseconds.
*RST
0.00000000E
000
Key Entry
Pulse Delay
:PULM:INTernal:FREQuency
Supported
All with Option UNW
[:SOURce]:PULM:INTernal:FREQuency <frequency>|MAXimum|MINimum|UP|DOWN
[:SOURce]:PULM:INTernal:FREQuency?
This command sets the pulse rate for the internally- generated square wave using the variable
<frequency>
. The command, used with the
UP|DOWN
parameters, will change the frequency by a
user- defined step value. Refer to the
:PULM:INTernal:FREQuency:STEP
command for setting the value
associated with the
UP
and
DOWN
choices.
This command is used when
SQUare
is the pulse modulation type. Refer to
:PULM:SOURce
command
for the pulse modulation type selection.
Example
:PULM:INT:FREQ 1MHz
The preceding example sets the square wave pulse rate to 1 megahertz.
*RST
4.00000000E
002
Range
0.1Hz–10MHz
Key Entry
Pulse Rate
:PULM:INTernal:FREQuency:STEP
Supported
All with Option UNW
[:SOURce]:PULM:INTernal:FREQuency:STEP[:INCRement] <freq>|MAXimum|MINimum|DEFault
[:SOURce]:PULM:INTernal:FREQuency:STEP[:INCRement]?
Range
Internal Free Run
: depends on pulse period and pulse width settings
Internal Triggered, Adjustable Doublet, & Triggered Doublet
: 70nS to (42 S - 10
nS - pulse width)