Agilent X-Series Signal Generators SCPI Command Reference
263
Arb Commands
Dmodulation Subsystem–N5172B/82B with Option 431 ([:SOURce]:RADio:DMODulation:ARB)
:PHASe:NOISe:TRACe?
Supported
N5172B/82B with Option 431 and 432
[:SOURce]:RADio:DMODulation:ARB:PHASe:NOISe:TRACe?
<startFreq>,<stopFreq>,<numSamples>
This query returns the theoretical phase noise amplitude mask applied with the current settings if
the phase noise feature is on. This mask does not take the natural phase noise of the instrument into
account, only the impairment from the phase noise feature. The output is over the start frequency to
the stop frequency for the number of samples specified. The samples are taken at logarithmic
frequency steps and the output is in dBc/Hz.
:RETRigger
Supported
N5172B/82B with Option 431
[:SOURce]:RADio:DMODulation:ARB:RETRigger ON|OFF|IMMediate
[:SOURce]:RADio:DMODulation:ARB:RETRigger?
This command enables or disables the ARB retriggering mode; the retrigger mode controls how the
retriggering function performs while a waveform is playing.
ON (1)
This choice specifies that if a trigger occurs while a waveform is playing, the
waveform will retrigger at the end of the current waveform sequence and play
once more.
OFF (0)
This choice specifies that if a trigger occurs while a waveform is playing, the
trigger will be ignored.
IMMediate
This choice specifies that if a trigger occurs while a waveform is playing, the
waveform will reset and replay from the start immediately upon receiving a
trigger.
*RST
ON
:SCLock:RATE
Supported
N5172B/82B with Option 431
[:SOURce]:RADio:DMODulation:ARB:SCLock:RATE <val>
[:SOURce]:RADio:DMODulation:ARB:SCLock:RATE?
This command sets the sample clock rate.
The variable <val> is expressed in units of Hertz (kHz
MHz)
*RST
+4.00006
Range
1E3 to 2E8
Key Entry
ARB Sample Clock
Range
<startFreq>
1 Hz to 100 MHz
<stopFreq>
1 Hz to 100 MHz
<numSamples>
1 to 8192
Key Entry
On
Off
Immediate