32
Agilent X-Series Signal Generators SCPI Command Reference
Basic Function Commands
Digital Modulation Subsystem—N5172B/82B ([:SOURce])
*RST
+0.00000
Range
25 to 25 mV
Key Entry
Diff. Mode Q Offset
Remarks
This command is effective only if the state of the I/Q adjustment function is set to
ON. Refer to the
:DM:IQADjustment[:STATe]
command.
:DM:IQADjustment:EXTernal:INPut:GAIN
Supported
N5172B/82B
[:SOURce]:DM:IQADjustment:EXTernal:INPut:GAIN <val>
[:SOURce]:DM:IQADjustment:EXTernal:INPut:GAIN?
This command adjusts the External Input I/Q Gain Balance.
The variable <val> is expressed in units of decibels (dB), and the minimum increment is 0.001dB.
*RST
+0.00000
Range
1 to 1
Key Entry
External Input I/Q Gain Balance
:DM:IQADjustment:EXTernal:IOFFset
Supported
N5172B/82B
[:SOURce]:DM:IQADjustment:EXTernal:IOFFset <value>
[:SOURce]:DM:IQADjustment:EXTernal:IOFFset?
This command sets the offset voltage for a signal applied to the external I Input connector.
The variable <value> is expressed in units of volts (mV–V).
*RST
+0.00000
Key Entry
External Input I Offset
Range
100 to 100 mV
Remarks
This command is effective only if the state of the I/Q adjustment function is set to
ON. Refer to the
:DM:IQADjustment[:STATe]
command.
:DM:IQADjustment:EXTernal:QOFFset
Supported
N5172B/82B
[:SOURce]:DM:IQADjustment:EXTernal:QOFFset <value>
[:SOURce]:DM:IQADjustment:EXTernal:QOFFset?
This command sets the offset voltage for a signal applied to the External Q Input connector.
The variable <value> is expressed in units of volts (mV–V).
*RST
+0.00000