178
Agilent X-Series Signal Generators SCPI Command Reference
System Commands
System Subsystem (:SYSTem)
:PON:TYPE
Supported
All
:SYSTem:PON:TYPE PRESet|LAST|USER
:SYSTem:PON:TYPE?
This command sets the defined conditions for the signal generator at power on.
PRESet
This choice sets the conditions to factory– or user–defined as determined by the
choice for the preset type.
LAST
This choice retains the settings at the time the signal generator was last powered
down.
USER
This choice sets the power on state to be the user preset value.
Key Entry
Power On Last Preset
Remarks
The setting enabled by this command is not affected by signal generator power–on,
preset, or
*RST
.
For a comparison of the SCPI preset commands, refer to
Table 3- 1, “Preset SCPI
Commands Overview,” on page 178
.
:PRESet
NOTE
If this SCPI command is not responding as expected, use the E4428C/38C compatibility
command:
:SYST:PRESet:TYPE:NORMal
to return the front- panel
Preset
key to its factory
default functionality.
Supported
All
SYSTem:PRESet
This command returns the signal generator to a set of defined conditions. It is equivalent to pressing
the front- panel
Preset
hardkey.
Key Entry
Preset
Remarks
The defined conditions are either factory– or user–defined.
For a comparison of the SCPI preset commands, refer to
Table 3- 1, “Preset SCPI
Commands Overview,” on page 178
.
Table 3-1
Preset SCPI Commands Overview
Command
Description
Remarks
*RST
This IEEE 488.2 Common Command uses the
factory preset settings for the instrument
preset.
Optimized for automated
testing
:SYSTem
:PRESet:PERSistent
Only the instrument’s persistent parameters
are returned to factory default value.