Service
Mini Med Series
0000061122
Troubleshooting
Processor Problems,
Continued
Symptom
Probable Cause
Remedy
Note: The interupter switch may be bypassed
for troubleshooting purposes only
by
either utilizing the hold down tool (P/N 0000021801) to keep the switch activated.
Jumping the pins on J-14 on the logic board is an alternate method.
5-5
11. Increase in image
12. Decrease in image
density.
13. Over replenishing.
14. No functions
A. Film is over exposed.
B. Developer temperature too high.
C. Excessive developing time due
to mechanical binding.
D. New developer improperly mixed.
.
A. Film is under exposed.
B. Developer under replenished
or exhausted.
C. Developer temperature too low.
D. Developer time too short.
E. New developer improperly
mixed.
A. Shorted film presence switch.
B. Shorted manual replenishment
switch.
C. Pendulum magnet weak or
stuck in the up position.
D. Logic failure
A. Interupter switch not activated.
B. Interupter switch defective.
A. Coordinate exposure
with developing time.
B. Have service technician
troubleshoot developer
temperature.
C. See symptom 1.
D. Dump and mix fresh,
following mfg's
instructions exactly.
A. Coordinate exposure
B. Change developer.
Check replenishment
rates.
C. See Symptom 3.
D. Check developer for
low level.
E. Dump developer and
mix fresh.
A. Replace switch.
B. Replace switch.
C. Clean or replace
pendulum.
D. Replace logic board.
A. Put top cover in place.
B. Replace switch.
Summary of Contents for Mini-Medical Series
Page 4: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 26: ...Introduction Mini Med Series 0000061122 ...
Page 28: ......
Page 30: ......
Page 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Page 46: ......
Page 56: ......
Page 66: ......
Page 98: ......
Page 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Page 122: ......
Page 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Page 130: ......
Page 131: ......