Operation
Mini Med Series
0000061122
User Controls,
Continued
2. Manual Replenishment Switch
Provides for manual operation of replenishment pumps. Use to “top off” tanks or to
turn over chemistry when activity levels have dropped.
3. Power ON LED
Lights when Power Switch is ON.
4. Dev Temp LED
Lights when developer heater is ON. Wait for light to cycle OFF before first use each
day.
5. Wait LED
Illuminates when Film Feed switch is activated. To prevent fogging of film, wait until
lamp goes OFF or beeper sounds before opening feed tray.
6. Low Dev LED
Lamp ON indicates that developer is too low for safe operation. To prevent damage to
the processor, the developer heat function is turned OFF when a low level condition
exists.
7 Drain Valves
Drains the Developer, Fixer, Wash tank and recirculation pumps.
8. Overflow Lines
Drain lines from developer, fix & wash overflow ports.
9 Top Cover Interlock Switch
To prevent accidental injury from moving parts, a safety switch is interlocked with the
processor’s top cover. If the top cover is removed, the processor automatically shuts
down. This interlock may be overridden for service use only by using the service hold
down tool included in the Maintenance Kit provided with each unit.
Caution: DO NOT attempt to process film in this unit when the interlock is overridden.
3-2
Summary of Contents for Mini-Medical Series
Page 4: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 26: ...Introduction Mini Med Series 0000061122 ...
Page 28: ......
Page 30: ......
Page 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Page 46: ......
Page 56: ......
Page 66: ......
Page 98: ......
Page 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Page 122: ......
Page 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Page 130: ......
Page 131: ......