Maintenance
Mini Med Series
0000061122
Prepare Fresh Chemicals
When the developer and fixer systems are clean, prepare and load fresh chemicals in accor-
dance with instructions in Section 3, Operation, and the manufacturer’s instructions and
precautions.
Monthly Lubrication
Refer to Figure 4-2, Maintenance Log and Figure 4-3, Lubrication Points and lubricate as
indicated. Be sure to clean off all old lubricants and any excessive new lubricants.
Annual Maintenance
Once a year, following a routine monthly cleaning, perform the following tasks on the
processor:
1) Inspect the drive gears on each rack and replace any gears that are excessively worn or
damaged.
2) Refer to Service Procedure 5-1. Inspect, and adjust or replace if necessary, the main
drive belt.
3) Refer to Service Procedure 5-2. Inspect and clean the fixer circulation pump. The
developer pump is usually cleaned adequately by systems cleaning and does not require
additional servicing.
4) Refer to Service Procedure 5-3. Inspect and clean the developer and fixer replenishment
pumps.
5) Refer to Figure 4-2, Maintenance Log and Figure 4-3, Lubrication Points and lubricate
as indicated.
Be sure to clean off all old lubricants and any excessive new lubricants.
4-7
Summary of Contents for Mini-Medical Series
Page 4: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 26: ...Introduction Mini Med Series 0000061122 ...
Page 28: ......
Page 30: ......
Page 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Page 46: ......
Page 56: ......
Page 66: ......
Page 98: ......
Page 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Page 122: ......
Page 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Page 130: ......
Page 131: ......