Installation
Mini Med Series
0000061122
Operational Checkout
Continued
Caution: Never attempt to operate the processor without liquid in the tanks.
6) Switch the Power Switch to ON.
7) The transport system will run at Process speed for the duration of one processing cycle,
the recirculation pumps will operate and the dryer heaters and fans will be activated.
NOTE: If either of the recirculation pumps do not prime, squeeze the rubber elbow on the
inlet side of the pump(s) to purge the air from the line.
8) Inspect all rack modules to verify that they are turning freely.
9) Carefully inspect the underside of the processor for any signs of leakage. Correct if
necessary.
10) Operate the Manual Replenishment switch to run the replenishment pumps until the
developer and fixer tanks are full of water to the overflow port.
11) Activate the Film Feed switch with a piece of film. The Wait light will turn ON and
every few seconds the replenishment pumps will cycle. Remove the film from the sensor.
In a few seconds the Wait light will go out and the beeper will sound.
12) When the Dev Temp lamp turns OFF, check the temperature of the developer with a
metal stem or digital thermometer*. If it is not correct for the film you will be using, adjust
the temperature as outlined in Section 5, Service.
*Warning:
Do not use a mercury thermometer. If a mercury thermometer breaks, it will
contaminate the machine
2-10
Summary of Contents for Mini-Medical Series
Page 4: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 26: ...Introduction Mini Med Series 0000061122 ...
Page 28: ......
Page 30: ......
Page 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Page 46: ......
Page 56: ......
Page 66: ......
Page 98: ......
Page 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Page 122: ......
Page 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Page 130: ......
Page 131: ......