Installation
Mini Med Series
0000061122
Transport Film
Transport several pieces of film of your usual size(s) through the processor. Inspect and, if
necessary, correct for the following:
1) Film Feed switch operation. The Wait lamp should stay on continuously until a few
seconds after the trailing edge of the material being processed is clear of the Film Feed
switch.
As the Wait light turns off, an audible beeper will sound indicating it is safe to feed in
another piece of material.
The processor will remain in the process mode for approximately 5 minutes after the film
feed switch is released.
2) Drift or Skewing. The film should feed through the processor in a straight line. If it
drifts, skews or wrinkles, check the racks for proper seating or loose assembly screws. Be
certain you are feeding the film in straight before checking racks.
3) Drying. Be sure the dryer is operating properly. Film processed in water alone may still
be slightly tacky or damp when exiting the processor.
Complete Checkout
1) Turn the Power Switch OFF. Unplug the power cord.
2) Drain each of the processing tanks and the replenishment containers for the developer
and fix replenishment systems. Close all drain valves.
3) Wipe any excess water from the racks and tanks.
The processor is now ready to be charged with fresh chemistry as instructed in Section 3,
OPERATION.
2-11
Summary of Contents for Mini-Medical Series
Page 4: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 26: ...Introduction Mini Med Series 0000061122 ...
Page 28: ......
Page 30: ......
Page 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Page 46: ......
Page 56: ......
Page 66: ......
Page 98: ......
Page 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Page 122: ......
Page 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Page 130: ......
Page 131: ......