Accessories
Mini Med Series
0000061122
Side Mount Drain Manifold Kit
The
Side Plumbing Kit
, p/n 9992305002, allows the re-
positioning of the front mounted plumbing manifold to
either right or left side of a Mini-Medical processor.
1) Following the dimensions on the illustration be-
low, drill four (4) 3/16" holes in the right or left
side of the processor for the manifold brackets.
2) Remove the original plumbing manifold from the
front of the processor (2 screws).
3) Carefully re-route the hoses to allow the plumbing
manifold to be relocated to the right or left side of
the processor. NOTE: It may be necessary to remove and
then reconnect one or more of the hoses from the man-
ifold during this process to obtain the best routing
without kinking the hoses.
4) Install the plumbing manifold into the manifold
bracket on the side of the processor selected. Check
for leaks wlth water or chemicals.
5) Install the filler plate in the front opening.
6) After charging the processor, carefully inspect all
connections for leaks and proper drainage.