![AFP Imaging Mini-Medical Series Installation, Operation & Service Parts Manual Download Page 29](http://html.mh-extra.com/html/afp-imaging/mini-medical-series/mini-medical-series_installation-operation-and-service-parts-manual_2864734029.webp)
Mini Med Series
0000061122
Attention:
When testing or operating the processor with water (as opposed to chemistryj), there will be a
LOW LEVEL condition in effect which will disable the solution heater. This is due to the fact that water
by itself cannot conduct well enough. To prevent this, add 1-2 tablespoons of salt or a cup of used or
fresh developer to the developer tank.