Operation
Mini Med Series
0000061122
NOTE: For operator convenience, the
LED Display Panel and the Power Switch
Panel may be switched, right to left,
during installation of the processor.
See Page 2-8 for details.
Figure 3-1, User Controls
Controls and Indicators
All of the user controls and indicators for operation of Mini-Medical Processors are
located on the front of the processor. These controls are described below and on the
following page.
User Controls
1. Power Switch
OFF
All power to processor is OFF.
ON
Processor is ON, (runs one approximate 5 minute process cycle initially),
then reverts to standby mode. Circulation pumps, developer heater and air
circulation fan are ON. Transport, replenishment system and dryer will
operate when Film Feed switch is activated. Power On lamp will light.
CENTER
This switch also serves as the circuit breaker for the processor. If tripped to
the center position reset to OFF, then turn ON. If the switch trips again, the
processor probably needs a service call. Do not use the processor if it trips
off repeatedly.
1
2
4
8
7
3
9
6
5
3-1
Summary of Contents for Mini-Medical Series
Page 4: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 26: ...Introduction Mini Med Series 0000061122 ...
Page 28: ......
Page 30: ......
Page 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Page 46: ......
Page 56: ......
Page 66: ......
Page 98: ......
Page 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Page 122: ......
Page 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Page 130: ......
Page 131: ......