Service
Mini Med Series
0000061122
Troubleshooting
Processor Problems,
Continued
Symptom
Probable Cause
Remedy
5-4
7. Films overlap or
become skewed
during transport.
8. Film is tacky, wet or
curled when leaving
dryer.
9. Dirt particles on film.
10. Scratches on film
emulsion.
A. Bound rollers.
B. Rack end plate bearings worn.
C. Missing rack springs.
A. Improper fixing.
B. Dryer temperature too low.
A. Foreign particles in dryer or on
squeegee rollers.
B. Algae deposits on film.
C. Foreign particles on squeegee
A. Dirt on feed rollers
B. Chemicals crystallized on
underside of top film guides.
C. Dirt or silver accumulation on
rollers.
D. Roller in rack not turning.
E. Dirty feed tray surface.
A. Clean each roller;
check for causes of
binding.
B. Replace bearings.
C. Replace springs.
A. See 5B.
B. Check dryer for
proper operation.
A. Run several outdated
sheets of unexposed
film.
B. Clean wash tanks &
racks with nylon
scrub brush and
warm water. Drain
wash tank each
night.
C. Clean squeegee
roller(s).
A. Clean feed rollers.
B. Clean film guides.
C. Clean rollers using a
nylon scrub pad and
warm water, or
developer systems
cleaner for developer
racks/fixer systems
cleaner for fix racks.
D. Check all rollers for
operation; repair as
required.
E. Clean feed tray.
Summary of Contents for Mini-Medical Series
Page 4: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 26: ...Introduction Mini Med Series 0000061122 ...
Page 28: ......
Page 30: ......
Page 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Page 46: ......
Page 56: ......
Page 66: ......
Page 98: ......
Page 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Page 122: ......
Page 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Page 130: ......
Page 131: ......