Maintenance
Mini Med Series
0000061122
Weekly Cleaning
1) Remove the developer, fixer and wash racks. To prevent chemical contamination, wash
off each rack separately with clean, lint-free cloths and warm water. Clean each roller over
its entire surface. Use isopropyl alcohol if necessary to remove traces of adhesives.
NOTE
: Soft scrub pads, such as nylon net over sponge, work well on rollers. Metallic, or
non-metallic, scrub pads such as "Scotchbrite", must not be used on rollers as they will
damage the roller surface.
2) Inspect each rack thoroughly. Verify that the rollers turn freely and that all guides and
baffles are properly in place. Carefully set each rack aside to drain and dry while you are
cleaning the rest of the processor.
3) Clean the outside surfaces of the processing tank, using warm water with a sponge or
non-metallic scrub pad.
Caution: Never use steel wool on any part of the processor as its residue may cause
rust to form on the metallic parts of the processor.
4) Clean the dryer rollers and rack parts with a damp cloth and wipe dry.
5) Replace all removed racks and other parts.
4-4
Summary of Contents for Mini-Medical Series
Page 4: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 26: ...Introduction Mini Med Series 0000061122 ...
Page 28: ......
Page 30: ......
Page 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Page 46: ......
Page 56: ......
Page 66: ......
Page 98: ......
Page 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Page 122: ......
Page 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Page 130: ......
Page 131: ......