![AFP Imaging Mini-Medical Series Installation, Operation & Service Parts Manual Download Page 24](http://html.mh-extra.com/html/afp-imaging/mini-medical-series/mini-medical-series_installation-operation-and-service-parts-manual_2864734024.webp)
Introduction
Mini Med Series
0000061122
General Specifications,
Continued
Environmental Conditions
Temperature:
40-80 Degrees f.
Humidlty:
40% - 60% RH.
Electrical Requirements
120 VAC, 15 amps, 60 Hz.
230 VAC, 7.5 amps, 50 Hz. (Optional)
Dimensions
Width 22" (56.1 Cm)
Stand Only:
Width 22.875"
Height: 24.5" (62.2 Cm)
Height 29.5"
Length: 33" (84.1 Cm)
Length 22.125"
(Including feed tray)
Allow 1" (approximately) for leveling
Weight
Approximate Shipping Wt.: 110 lbs.
Approximate Operating Wt.: 160 lbs.
Air Conditioning Heat Load (Approximate)
Total Heat @ 60Hz
Process Mode
Standby Mode
2800 B.T.U./Hour
1800 B.T.U./Hour
Darkroom venting is required. Use a blower and vent combination that allows for approxi-
mately 300 CFM air flow through the darkroom.
Component Power Requirements
Component
Amperage @ 115VAC
Solenoid
0.10
Developer Heater
4.55
Recirculation Pumps (2)
0.22 Ea.
Replenisher Pumps (2)
0.35 Ea.
Dryer Lamps (2)
2.17 Ea.
Fan Motors (2)
0.44 Ea.
Drive Motor
0.76
Total:
11.77 Amps
1-8
Summary of Contents for Mini-Medical Series
Page 4: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 26: ...Introduction Mini Med Series 0000061122 ...
Page 28: ......
Page 30: ......
Page 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Page 46: ......
Page 56: ......
Page 66: ......
Page 98: ......
Page 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Page 122: ......
Page 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Page 130: ......
Page 131: ......