Installation
Mini Med Series
0000061122
Connect Plumbing
Wash water into the processor is controlled by the water solenoid valve. When the proces-
sor is in the process mode the solenoid is actuated, allowing wash water to flow into the
bottom of the wash tank. Excess water in the wash tank flows over the overflow port and
down the drain.
To connect the wash water proceed as follows: (See Figures 2-2 and 2-3)
1) Route the supplied reinforced water hose from the output of your water panel to the
fitting on the wash water solenoid under the right side of the processor.
2) Route the clear wash water drain line and the clear overflow line from the utility section
of the processor to the building drain or a suitable overflow container.
CAUTION: If you are draining your processor directly into a sanitary sewer, be certain that
such connections are in accordance with local plumbing codes and comply with all local
and federal EPA anti-pollution requirements.
DO NOT drain the processor into any drain lines that are made of copper pipe as chemical
reactions will quickly damage the pipes.
"No Plumbing" System Option
The available “No Plumbing” wash water recirculation system, P/N 9992305003 (115 VAC)
or 9992305004 (230), allows the installation of any of the Mini-Medical Series processors
without the need for an external water supply or drains. If this system is being installed
with the processor, refer to the instructions packed with that unit and to Section 7, Accesso-
ries, in this manual.
2-7
Summary of Contents for Mini-Medical Series
Page 4: ......
Page 10: ......
Page 12: ......
Page 14: ......
Page 16: ......
Page 26: ...Introduction Mini Med Series 0000061122 ...
Page 28: ......
Page 30: ......
Page 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Page 46: ......
Page 56: ......
Page 66: ......
Page 98: ......
Page 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Page 122: ......
Page 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Page 130: ......
Page 131: ......