R3132 Series Spectrum Analyzer Operation Manual
A.1 ERROR MESSAGE
A-2
Error No.
Error Message
Description
15
Not available.
Please input External 10MHz Ref.
Cannot set to EXT because 10 MHz external refer-
ence signal is not input.
Please input the external reference signal.
16
External 10MHz Ref. is not detected.
10MHz Ref. changed to Internal.
Since 10 MHz external reference signal has not
been detected, the reference signal source is
switched to INT.
Please check the external reference signal.
17
Not available.
Trigger source is Free Run or Line.
Trigger slope (TV modulation polarity) cannot be
switched b and - because the trigger
source is set to Free Run or Line.
18
Gated sweep setup mode.
Please select the same Gate source.
A trigger source can be selected only from the
Gate sources.
19
Not available.
RBW is 1MHz.
Cannot change to QP detection mode because an
RBW of 1MHz is selected for the PEAK detection
mode.
20
Not available.
EMC Detector is QP.
RBW cannot be set to 1 MHz because QP detec-
tion mode is selected.
21
Scale is not Linear mode.
Please select Linear scale.
[Level
→
Linear]
Vertical scale is not in Linear mode.
Please change it to Linear mode.
22
Not available.
Channel tables are all disabled.
Channels cannot be set because all channel tables
are disabled.
23
Not available.
Channel table contains no data.
Channels cannot be set because all channel tables
have no data.
24
Peak detector is active.
Please change to Normal.
[EMC
→
Detector mode]
Peak detector mode is turned on.
Turn Normal mode on.
30
Parameter is out of range.
The value for the ACP CS/BS table is out of range.
Please change the span which can be measured.
31
Incorrect data.
Set span to (1.0 + alpha) * Tf or more.
The value for
√
Nyquist filter is out of range.
Change the measurement condition to meet the
following:
Frequency span > (1.0 + Rolloff factor)
×
Symbol
rate
32
Frequency table contains no data.
Cannot execute the ACP function because the
ACP CS/BC table has no data.
33
Editor is active.
Please quit the editor first.
The editor mode is disabled while this operation is
in process.
Quit the editor mode first.
34
Spurious table contains no data.
Enter data into the spurious table because it has no
data.