R3132 Series Spectrum Analyzer Operation Manual
5.2.16 Frequency Response
5-64
30. On the R3132 series, press FREQ and
∆
to increase center frequency by 100
MHz.
31. On the signal generator, increase the frequency of output by 100 MHz.
32. On the power meter, increase correction frequency by 100 MHz.
33. Repeat step 28 through 32 for every center frequency by 100 MHz step up the
center frequency to 3.2 GHz (for R3132 up to 3 GHz, For R3132N up to 2.2 GHz)
listed in the performance verification record sheet.
Measuring frequency response in the frequency range: 3.2 GHz to 8 GHz for R3162
Measuring frequency response in the frequency range: 3.3 GHz to 7.1 GHz for R3172/82
(Note: For testing this frequency range, preselector tune is required)
34. For setting the measurement conditions, perform step 1 through 9.
35. Press MKR, Peak Menu and Cont Peak ON/OFF(ON) to set continuous peak
search mode.
36. On the R3162/72/82, set center frequency to 3.3 GHz.
37. On the signal generator, set the frequency to 3.3 GHz.
38. On the power meter, set the correction frequency to 3.3 GHz.
39. On the R3162/72/82, press as follows to tune the preselector.
FREQ, 1/2_more, Presel, Auto Tune
40. After the auto tuning has completed, adjust signal generator output level so that
the marker reading is -10 dBm
±
0.09 dB.
41. Record the display of power meter reading with reverse sign in the performance
verification record sheet.
42. On the R3162/72/82, press FREQ and
∆
to increase center frequency by 100
MHz.
43. On the signal generator, increase the frequency of output by 100 MHz.
44. On the power meter, increase the correction frequency by 100 MHz.
45. Repeat steps 39 through 44 for every center frequency by 100 MHz step up the
center frequency to 8 GHz (for R3162) or 7 GHz (for R3172/82) listed in the per-
formance verification record sheet.