R3132 Series Spectrum Analyzer Operation Manual
4.2.9 GPIB Command Codes
4-40
Table 4-14Power (3 of 3)
(*1) n=Number of measurement points (0 thru 15)
m=Number of spurious signals (0 thru 10)
f=Spurious frequency
l=Spurious level
j=Judgment result (0: Pass, 1: Fail)
<DLM>=Delimiter
(*2) (*3) (*4)Perform the corresponding operation when the spurious measurement mode is turned on.
(*3) Specify the asterisk in order of the start frequency, stop frequency, RBW, sweep time and limit level.
(*4) Specify the asterisk in order of the center frequency, RBW, sweep time and limit level.
Function
Listener Code
Talker Request
Code
Output Format
Result display mode REL
ABS
SPMMOD REL
SPMMOD ABS
SPMMOD?
0:REL
1:ABS
Spurious measurement
execution Freq
Time
Spurious measurement OFF
SPURI FREQ
SPURI TIME
SPURI OFF
SPURION?
0:OFF
1:Freq
2:Time
Spurious measurement result
–––
SPURI?
n<DLM>,
m1<DLM>,f1,l1,j1<DLM>,
…
,
fm1,lm1,jm1<DLM>,
m2<DLM>,f1,l1,j1<DLM>,
…
,
fm2,lm2,jm2<DLM>,
: : :
mn<DLM>,f1,l1,j1<DLM>,
…
,
fmn,lmn,jmn<DLM>(*1)
Table selection
SPRTBL*
SPRTBL?
Integer(*2)
Table input
Freq
Time
SPRIN*(*3)
SPRFIN*(*3)
SPRTIN*(*4)
–––
–––
–––
–––
–––
–––
Table deletion
SPRDEL(*2)
–––
–––
Sweep count specification
ON
OFF
SPRCNT [ON,]*
SPRCNT OFF
SPRCNT?
SPRCNTON?
Integer
0:OFF
1:ON
Pass/fail judgment value
LOW
UP
SPRJ LOW
SPRJ UP
SPRJ?
0:LOW
1:UP
Single Measure
ON
OFF
SIMS ON
SIMS OFF
SIMS?
0: OFF
1: ON