R3132 Series Spectrum Analyzer Operation Manual
5.2.10 Scale Fidelity
5-39
Measuring the Linear Scale Fidelity
23. On the signal generator, set controls as follows:
Frequency:
11 MHz
Output Level:
0 dBm
24. On the external attenuators, set attenuation 0 dB.
25. On the R3132 series, after preset, set controls as follows:
Center Frequency:
11 MHz
Frequency Span:
10 kHz
Reference Level:
0 dBm
RBW:
3 kHz
VBW:
1 kHz
Input Attenuator:
20 dB
26. On the R3132 series, press as follows to set vertical display mode to linear.
LEVEL, Linear
27. On the R3132 series, press as follows to set continuous peak search mode.
MKR, Peak Menu, Cont Peak ON/OFF(ON)
28. On the signal generator, precisely set output level to the R3132 series reference
level while reading the marker level on the screen.
29. On the R3132 series, press SINGLE for single sweep.
30. Read the level value displayed on the signal generator and set the value as the ref-
erence value (Ref.).
31. Then set level of signal generator level to the 0.92 dB lower than the reference
value.
32. On the R3132 series, press SINGLE for single sweep.
33. Read the marker level and record it in the performance verification record sheet.
34. Repeat steps 28 through 34 for each value listed in Table 5-16.