R3132 Series Spectrum Analyzer Operation Manual
5.2.19 Gain Compression
5-77
5.2.19 Gain Compression
This test measures the analyzer’s gain compression using two signals that are 1 MHz apart. First the test
places a -30 dBm signal at the input of the R3132 series (the R3132 series reference level is also set to -30
dBm).
Then the specified signal level is input to the R3132 series, overdriving its input. The decrease in the first
signal’s amplitude (gain compression) caused by the second signal is the measured gain compression.
Specification:
Apply for R3132
> -0 dBm (mixer input level)
Preamplifier OFF, Frequency Range 200 MHz to 3 GHz
> -25 dBm (RF input level)
Preamplifier ON, Frequency Range 200 MHz to 3 GHz
Apply for R3132N
> +107 dB
µ
V (mixer input level) Preamplifier OFF, Frequency Range 200 MHz to 2.2 GHz
> +82 dB
µ
V (RF input level)
Preamplifier ON, Frequency Range 200 MHz to 2.2 GHz
Apply for R3162
> -0 dBm (mixer input level)
Preamplifier OFF, Frequency Range 200 MHz to 8 GHz
> -25 dBm (RF input level)
Preamplifier ON, Frequency Range 200 MHz to 3.3 GHz
Apply for R3172
> -0 dBm (mixer input level)
Preamplifier OFF, Frequency Range 200 MHz to 3.3 GHz
> -5 dBm (mixer input level)
Preamplifier OFF, Frequency Range 3.2 MHz to 26.5 GHz
> -25 dBm (RF input level)
Preamplifier ON, Frequency Range 200 MHz to 3.3 GHz
Apply for R3182
> -0 dBm (mixer input level)
Preamplifier OFF, Frequency Range 200 MHz to 3.3 GHz
> -5 dBm (mixer input level)
Preamplifier OFF, Frequency Range 3.2 MHz to 40 GHz
> -25 dBm (RF input level)
Preamplifier ON, Frequency Range 200 MHz to 3.3 GHz