R3132 Series Spectrum Analyzer Operation Manual
6.4.1 External Mixer Performance Verification Test Procedures
6-59
25. Record the result in the actual value column on the performance verification test
data sheets.
26. Repeat steps 19 through 25 for the frequency setting both of R3172/R3182 center
frequency and SG output frequency up-to 110 GHz step by 100 MHz.