R3132 Series Spectrum Analyzer Operation Manual
5.2.21 Residual Response
5-86
5.2.21 Residual Response
This test checks for residual responses under preamplifier on and off cases.
Any response located above the display line is measured in a narrow frequency span and RBW.
The RF INPUT is terminated.
Specification:
Apply for R3132
≤
-100 dBm Preamplifier OFF, Frequency Range 1 MHz to 3 GHz
≤
-105 dBm Preamplifier ON, Frequency Range 1 MHz to 3 GHz
Apply for R3132N
≤
+11 dB
µ
V Preamplifier OFF, Frequency Range 1 MHz to 2.2 GHz
≤
+6 dB
µ
V Preamplifier ON, Frequency Range 1 MHz to 2.2 GHz
Apply for R3162
≤
-100 dBm Preamplifier OFF, Frequency Range 1 MHz to 3.3 GHz
≤
-90 dBm Preamplifier OFF, Frequency Range 3.2 GHz to 8 GHz
≤
-105 dBm Preamplifier ON, Frequency Range 1 MHz to 3.3 GHz
Apply for R3172
≤
-100 dBm Preamplifier OFF, Frequency Range 1 MHz to 3.3 GHz
≤
-90 dBm Preamplifier OFF, Frequency Range 3.2 GHz to 26.5 GHz
≤
-105 dBm Preamplifier ON, Frequency Range 1 MHz to 3.3 GHz
Apply for R3182
≤
-100 dBm Preamplifier OFF, Frequency Range 1 MHz to 3.3 GHz
≤
-90 dBm Preamplifier OFF, Frequency Range 3.2 GHz to 40 GHz
≤
-105 dBm Preamplifier ON, Frequency Range 1 MHz to 3.3 GHz
Instruments Required:
Instruments
Qty.
Recommended Model
Terminator 50
Ω
(for R3132/3162)
1
RNA
Terminator 50
Ω
(for R3172)
1
Terminator 50
Ω
(for R3182)
1
Terminator 75
Ω
(for R3132N)
1