R3132 Series Spectrum Analyzer Operation Manual
6.4 External Mixer OPT19
6-52
6.4
External Mixer OPT19
(1)
Introduction
This chapter provides OPT3172/3182+19 external mixer performance verification test procedures,
item by item as listed on Table 6-7.
Performance verification will be carried out under following condition.
Temperature range: 20
°
C to 30
°
C
Relative Humidity: 85 % or less
Table 6-7 List of Performance Verification Test Items
(2)
Test Equipment
Table 6-8 lists equipment for the performance verification test.
The equipment needed to perform all of the performance test.
Equipment lists for individual tests are provided in each performance verification test procedure.
Table 6-8 Equipment List for Performance Verification Test
No.
Test Description
Applicable Models
1
Noise Level
OPT3172/3182+19
2
Frequency Response
OPT3172/3182+19
No.
Description
Critical Specifications
for substitute Model
Model
Recommended
Manufacturer
Notes
1
Sweeper
No substitute model
recommended
83640
Agilent
Technologies
SG
2
Soure Module
No substitute model
recommended
83558A
Agilent
Technologies
-
3
10 dB Fixed Attenuator
No substitute model
recommended
521W
MPI
-
4
RF Power Sensor
No substitute model
recommended
W8486A
Agilent
Technologies
-
5
RF Power Meter
No substitute model
recommended
437B
Agilent
Technologies
-
6
RF Cable
Type: SMA(m)-SMA(m)
DCP-FF00092X02
Advantest
-
7
Spectrum Analyzer
No substitute model
recommended
R3172/R3182
Advantest
-