R3132 Series Spectrum Analyzer Operation Manual
5.2.16 Frequency Response
5-63
16. On the signal generator, increase the frequency of output by 100 MHz.
17. On the power meter, increase the correction frequency by 100 MHz.
18. Repeat steps 12 through 17 for every center frequency by 100 MHz step up the
center frequency to 3.2 GHz (for R3132, up to 3 GHz, for R3132N, up to 2.2
GHz) listed in the performance verification record sheet.
Under preamplifier on condition
19. On the power meter, Set power meter into dBm mode.
20. On the power meter, set the correction frequency to 30 MHz.
21. On the signal generator, set controls as follows:
Frequency:
30 MHz
Frequency Step:
100 MHz
Output Level:
-14 dBm
22. On the R3132 series, after preset set controls as follows:
Center Frequency:
30 MHz
CF Step:
100 MHz
Frequency Span:
40 MHz
Hi Sense:
ON
Input Attenuator:
10 dB
Reference Level:
-15 dBm
dB/div:
1 dB/div
RBW:
3 MHz
VBW:
1 kHz
23. Press MKR, Peak Menu and Cont Peak ON/OFF(ON) to set continuous peak
search mode.
24. On the signal generator, adjust output level so that the reading of peak search
marker is -20 dBm
±
0.09 dB.
25. On the power meter, set relative measurement mode.
26. On the signal generator, set frequency to 100 MHz.
27. On the power meter, set the correction frequency for 100 MHz.
28. On the signal generator, adjust output level so that reading of peak search marker
is -20 dBm
±
0.09 dB.
29. Record the display of power meter reading with reverse sign in the performance
verification record sheet.