R3132 Series Spectrum Analyzer Operation Manual
5.5 Performance Verification Record Sheet
5-150
(19) Sweep Time Accuracy
(20) Residual Response
•
For the R3132
•
For the R3132N
•
For the R3162/72/82
Sweep Time
Setting (s)
Test Data (s)
Specification
Result
Min. (s)
Measured
Value (s)
Max. (s)
Pass/Fail
50
45
44.1
45.9
5
4.5
4.41
4.59
500 m
450 m
441 m
459
50 m
45 m
44.1 m
45.9 m
5 m (Option)
4.5 m
4.41 m
4.59 m
500
µ
(Option)
450
µ
441
µ
459
µ
50
µ
(Option)
45
µ
44.1
µ
45.9
µ
Preamplifier
Frequency Range
Specification
Result
Min. (dBm)
Measured
Value
Max. (dBm)
Pass/Fail
OFF
1 MHz to 3 GHz
N/A
-100
ON
1 MHz to 3 GHz
N/A
-105
Preamplifier
Frequency Range
Specification
Result
Min (dB
µ
V)
Measured
Value
Max. (dB
µ
V)
Pass/Fail
OFF
1 MHz to 2.2 GHz
N/A
11
ON
1 MHz to 2.2 GHz
N/A
6
Preamplifier
Frequency Range
Specification
Result
Min. (dBm)
Measured
Value
Max. (dBm)
Pass/Fail
OFF
1 MHz to 3.3 GHz
N/A
-100
ON
1 MHz to 3.3 GHz
N/A
-105
OFF
3.2 MHz to 8 GHz
N/A
-90