R3132 Series Spectrum Analyzer Operation Manual
5.2.16 Frequency Response
5-62
CAUTION:
Use only 75
Ω
Ω
Ω
Ω
cables, connectors, or adapters on R3132N, or damage
to the input connector will occur.
5.
On the signal generator, set controls as follows:
Frequency:
30 MHz
Frequency Step:
100 MHz
Output Level:
-4 dBm
6.
On the R3132 series, after preset set controls as follows:
Center Frequency:
30 MHz
CF Step:
100 MHz
Frequency Span:
40 MHz
Reference Level:
-5 dBm
Input Attenuator:
10 dB
dB/div:
1 dB/div
RBW:
3 MHz
VBW:
1 kHz
7.
Press MKR, Peak Menu and Cont Peak ON/OFF(ON) to set continuous peak
search mode.
8.
On the signal generator, adjust output level so that reading of peak search marker
is -10 dBm
±
0.09 dB.
9.
On the power meter, set relative measurement mode.
Measuring frequency response in the frequency range: 9 kHz to 3.3 GHz for R3162/72/82
Measuring frequency response in the frequency range: 9 kHz to 3 GHz for R3132
Measuring frequency response in the frequency range: 9 kHz to 2.2 GHz for R3132N
Under preamplifier off condition
10. On the signal generator, set the frequency to 100 MHz.
11. On the R3132 series, set center frequency to 100 MHz.
12. On the power meter, set the correction frequency to 100 MHz.
13. Adjust output level of signal generator of peak search marker is -10 dBm
±
0.09
dB.
14. Record the display of power meter reading with reverse sign in the performance
verification record sheet.
15. On the R3132 series, press FREQ and
∆
to increase center frequency by 100
MHz step.