Operating Instructions — Analyze and Process Data
4500 Series of Instruments
System User Guide
82 of 138
RUO-IDV-05-0264-A
Show TICs from a Spectrum
•
Click
Explore > Show > Show TIC
.
The TIC opens in a new pane.
Show a Spectrum from a TIC
1. In a pane containing a TIC, select a range.
2. Click
Explore > Show > Show Spectrum
.
The spectrum opens in a new pane.
Generate XICs
XICs can be generated only from single period, single experiment chromatograms or spectra. To
obtain an XIC from multi-period or multi-experiment data, split the data into separate panes by
clicking the triangle under the x-axis. For more information about using the available icons, refer
.
Several methods are available for extracting ions to generate an XIC, depending on whether
chromatographic or spectral data is used.
contains a summary of methods that can be
used with chromatograms and spectra.
Prerequisite Procedure
•
. Open a data file containing spectra.
Tip!
To see an example data file, make sure that the Example project is selected.
Open the LIT folder, and then open the Reserpine.wiff file.
Tip!
Right-click inside a pane containing a spectrum and then click
Show
TIC
.
Tip!
Double-click in the TIC pane at a particular time to show the
spectrum.
Table 10-5 Summary of XIC Generation Methods
Method
Use with
Chromatogram
Use with
Spectrum
Extraction
Selected range
No
Yes
Extracts ions from a selected area in a
spectrum.
Maximum
No
Yes
Extracts ions from a selected area in a
spectrum using the most intense peak
in the selected area. This option
creates an XIC using the maximum
mass from the selected spectral range.
Summary of Contents for QTRAP 4500 LC/MS/MS system
Page 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Page 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Page 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Page 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...