Plumb the Diverter Valve
Plumb the diverter valve for both injector and diverter modes.
Plumb the Diverter Valve in Injector Mode
The diverter valve is a two position, 6 port valve. If you put the valve in Position A (
the sample flows through the external loop. When you switch the valve to Position B
), the sample is injected.
Figure 4-5 Diverter Valve—Injector Mode Position A
Item Description
1
Sample in
2
Waste out
3
Sample loop
4
Mobile phase in
5
To column
1
4
6
5
2
3
1
2
3
5
4