Operating Instructions — Batches
4500 Series of Instruments
System User Guide
66 of 138
RUO-IDV-05-0264-A
•
Select Vial Positions using the Locations Tab (Optional) on page 67
Submit a Sample or Set of Samples
1. In the
Batch Editor
, click the
Submit
tab.
2. If the
Submit Status
section contains a message about the status of the batch, do
one of the following:
•
If the message indicates that the batch is ready for submission, proceed to
•
If the message indicates that the batch is not ready for submission, make the
changes as indicated by the message.
3. Click
Submit
.
4. Save the file.
Change Sample Order
The order of the samples can be edited before they are submitted to the Queue.
•
On the
Submit
tab, double-click any of the numbers on the far left of the table (a very
faint square box is visible), and then drag them to the new location
Acquire Data
The system must not be in Tune mode when sample acquisition is started. Also, if the system
has been previously run that day and has not yet been set to Standby, sample acquisition will
start automatically.
1. In the navigation bar, click
Acquire
.
2. Click
View >
Sample Queue
.
The Queue Manager opens with all submitted samples.
Figure 9-3 Queue Manager
Item Description
1
The
Reserve Instrument for Tuning
icon should not be
pressed in.
3
1
2
Summary of Contents for QTRAP 4500 LC/MS/MS system
Page 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Page 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Page 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Page 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...