Operating Instructions — Analyze and Process Quantitative Data
4500 Series of Instruments
System User Guide
104 of 138
RUO-IDV-05-0264-A
10. Click
Show or Hide Parameters
twice.
11. Click
Specify Parameters - MQ III
.
12. Change the
Noise Percent
value.
13. Click
Apply
.
The peak is integrated closer to the baseline.
14. If the change does not improve the peak integration, then adjust the
Noise Percent
parameter until the optimal value is found.
15. To update the algorithm for all peaks, right-click in the pane and then click
Update
Method
.
Figure 11-6 Peak Review Pane
Item Description
1
Arrows: Click to move through the peaks.
2
Show or Hide Parameters: Click to show the integration parameters.
3
Integration parameters: Click to change the parameters.
Note:
The
Update Method
function only updates the algorithm values for
that specific analyte (or internal standard) and not all analytes.
1
2
3