Operating Instructions — Acquisition
Methods
An acquisition method consists of experiments and periods. Use the
Acquisition Method
editor
to create a sequence of periods and experiments for the instrument and devices.
Create Acquisition Methods
Only devices configured in the active hardware profile appear in the
Acquisition method
pane.
Any devices added to the hardware profile must also be added to existing acquisition methods.
For more information about devices, refer to the
Peripheral Devices Setup Guide.
1. In the navigation bar, under
Acquire
, double-click
Build Acquisition Method
.
2. In the
Acquisition Method Properties
tab, select a
Synchronization Mode
.
3. (Optional) Select the
Auto-Equilibration
check box and then type the desired
equilibration time, in minutes.
4. In the
Acquisition method
pane, click the
Mass Spec
icon.
5. On the
MS
tab, select a scan type.
6. Type values in the fields as required. Refer to
7. On the
Advanced MS
tab, type values in the fields as required. Refer to
.
8. Click a device icon and then select the parameters for the device. To configure the
syringe pump, refer to
Configure the Syringe Pump on page 53
9. Add any additional periods and experiments. Refer to
10. Click
File
>
Save
.
Configure the Syringe Pump
1. In the
Acquisition method
pane, click the
Syringe Pump
icon.
The
Syringe Pump Method Properties
tab opens in the
Acquisition Method
editor pane.
2. In the
Syringe Diameter (mm)
field, type the syringe diameter.
3. In the
Flow Rate
field, type the flow rate.
4. In the
Unit
list, select the units of flow.
Add an Experiment
1. Right-click the period and then click
Add experiment
.
An experiment is added below the last experiment in the period.
2. In the
Acquisition Method Editor
pane, select the appropriate device or instrument
parameters.
Summary of Contents for QTRAP 4500 LC/MS/MS system
Page 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Page 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Page 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Page 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...