13
System User Guide
4500 Series of Instruments
RUO-IDV-05-0264-A
119 of 138
Service and Maintenance Information
Regularly clean and maintain the system for optimal performance. For information on tuning
frequency, refer to
. For information on maintaining the ion source, refer to the ion
source
Operator Guide
.
provides a recommended schedule for cleaning and maintaining the system. Contact
a Qualified Maintenance Person (QMP) to order consumable parts. Contact an AB SCIEX
representative for maintenance service and support.
WARNING! Potential Radiation Hazard, Biohazard, or Toxic Chemical
Hazard: Determine whether instrument decontamination is required prior
to cleaning. Decontamination should be conducted prior to cleaning if
radioactive materials, biological agents, or toxic chemicals have been
used with an instrument.
Table 13-1 Maintenance Tasks
Component
Frequency
Task
For more information...
Curtain plate
Daily
Clean
Refer to
.
Orifice plate
(front)
Daily
Clean
Refer to
Clean the Front of the Orifice
.
Orifice plate
(front and rear)
As needed
Clean
Contact the local QMP or AB SCIEX Field
Service Employee (FSE).
QJet
®
ion guide
As needed
Clean
Contact the local QMP or FSE.
Q0 and IQ1 lens As needed
Clean
Contact the local QMP or FSE.
Instrument
surfaces
As needed
Clean
Refer to
Clean the Surfaces on page 120
Drain bottle
As needed
Empty
Refer to
Empty the Source Exhaust Drain
Roughing pump
oil
As needed
Check and fill
Contact the local QMP or FSE.
Instrument air
filter
Every 6
months
Inspect and
clean or replace
Contact the local QMP or FSE.
Roughing pumps
oil
Leybold
Sogevac
SV28: Every 6
months
Varian MS 40:
Annually
Replace
Contact the local QMP or FSE.
Electrode
As needed
Inspect and
clean or replace
Turbo V ™
Ion Source Operator Guide
.
Corona
discharge needle
As needed
Replace
Turbo V
™
Ion Source Operator Guide
.
Summary of Contents for QTRAP 4500 LC/MS/MS system
Page 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Page 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Page 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Page 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...