Operating Instructions — Tune and Calibrate
4500 Series of Instruments
System User Guide
50 of 138
RUO-IDV-05-0264-A
8. Select a
Tuning Solution
.
Depending on the solution selected, different modes are available.
i.
Click a polarity.
ii. If available, click
Q1
and
Q3
in the
Quad
section. If available, click the
required scan speeds.
iii. If available, click the scan speeds in the
LIT
section.
iv. If available, click
Excitation
in the MS/MS/MS section.
9. Click
Next
.
10. If the
Select a mode
page opens, select
Automatic
.
11. Click
Next
.
12. Click
GO
.
The
Verifying or Adjusting Performance
screen opens. After the process has
completed, the
Results Summary
opens. For more information, refer to the Help.
13. If applicable (depending on the options selected), change solutions when prompted.
Verify or Adjust Performance
The top left corner shows the part of the instrument that is being tuned.
Current Spectrum
: This graph shows the spectrum of the current scan, the optimal scan
selected by the software, or the scan at the current parameter value when the software results
are viewed in interactive mode.
The
Instrument Optimization Decision Plots
, in the top right graph, dynamically show the
intensity versus voltage curves of the parameters that are currently being optimized.
Results Summary
The
Results Summary
is a record of any instrument settings changes that were made by the
Instrument Optimization software. Refer to the
Results Summary
. This includes the location of data files and instrument settings backups, as well as step-
by-step changes and results during optimization.The
Results Summary
also shows a
verification report. This report contains a snapshot of the mass spectrum for each relevant mass
for the scan modes being verified. The spectrum is labelled with the target mass, where the mass
was found, mass shift, peak width, and peak intensity. The spectrum can be used as a visual
record of peak shape or scan mode performance. A summary table of results follows the spectra.
The
Results Summary
is saved as a document in the folder indicated at the top of the report.
Users can print the
Results Summary
or open a previously saved
Results Summary
.
Summary of Contents for QTRAP 4500 LC/MS/MS system
Page 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Page 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Page 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Page 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...