4500 Series of Instruments
System User Guide
134 of 138
RUO-IDV-05-0264-A
Table A-2
QTRAP 4500 System Parameters for LIT Scan Types Only
Access ID
Positive Ion Mode
Negative Ion Mode
CUR
(6)
20
10 to 55
20
10 to 55
CUR
(1)(2)(3)(4)(5)
20
10 to 55
20
10 to 55
CAD
High
Low; Medium; High
High
Low; Medium; High
IS
(1)(2)
5500
0 to 5500
-4500
-4500 to 0
IS
(5)
1500
0 to 2500
-1500
-2500 to 0
IS
(6)
1000
0 to 4000
-1000
-4000 to 0
NC
(3)(4)
3
0 to 5
–3
–5 to 0
TEM
(2)(3)(4)(5)
0
0 to 750
0
0 to 750
DP
100
0 to 300
–100
–300 to 0
EP
10
2 to 15
–10
–15 to –2
AF2
0.100
0 or 1
0.100
0 or 1
AF3
Mass-Speed Dependent
0 to 10
Mass-Speed Dependent
0 to 10
EXB
Mass-Speed Dependent
–165 to 0
Mass-Speed Dependent
0 to 165
CEM
1800
0 to 3300
1800
0 to 3300
GS1
20
0 to 90
20
0 to 90
GS2
0
0 to 90
0
0 to 90
CES
0
0 to 50
0
0 to 50