Chapter 8 Monitoring Non-Invasive Blood Pressure (NIBP)
8-14
www.zoll.com
9650-001355-01 Rev. M
Enable/Disable Smartcuf Motion Artifact Filter
Enabling the use of the Smartcuf motion artifact filter increases the accuracy of NIBP
measurements when motion artifact or weak pulses are present.
NIBP measurements can be adversely affected by many factors, such as cardiac arrthythmias,
sudden changes in blood pressure, body motions such as shivering and convulsions, bumping
the cuff, vibration, vehicle motion, or weak pulses. The Smartcuf filter synchronizes the NIBP
reading with the R-wave of the patient’s ECG to eliminate noise created by patient motion or
vibration.
Note:
The X Series unit must perform ECG monitoring when using Smartcuf.
If artifact is severe, a special symbol appears on the display and printouts:
Figure 8-5 High Artifact Symbol
By default, Smartcuf is
Enabled
.
There are situations where you may choose to disable Smartcuf. These may include
•
Very extreme motion artifact
•
Certain types of arrhythmias
•
Situations that prevent getting a ECG signal.
NIBP measurements can still be performed with Smartcuf disabled.
Start/Stop TurboCuf
Selecting
Start TurboCuf
starts Short-term Automatic (STAT) NIBP measurements. The X
Series unit begins its first NIBP measurement, after which it continues to perform as many
NIBP measurements as possible over a 5-minute period.
Select
Stop TurboCuf
to immediately stop STAT measurements.
WARNING! Repeated use of STAT measurements on the same patient over a short time interval can
affect blood pressure readings, limit circulation to the limb, and cause injury to the
patient.
Setting the NIBP Display Format
You can specify the appearance of the NIBP numeric display format. The display format allows
you to choose whether or not you will display MAP measurement (
M
) with the
Systolic (
S
) and Diastolic (
D
) measurements in one of the following formats:
•
S/D
•
S/D (M)
(default display)
•
(M) S/D
Note:
If a MAP alarm is active when the
S/D
format is selected, the display format will be
S/D (M)
until you clear the MAP alarm.
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...