Chapter 9 Monitoring CO2
9-10
www.zoll.com
9650-001355-01 Rev. M
Setting Upper and Lower Respiratory Rate (RR/BR) Limits
Initially, the RR/BR Alarm menu specifies that RR/BR
alarms are enabled (
ON
) or disabled
(
OFF
), and displays the default Upper and Lower respiratory limits. The upper and lower limits
can be
ON
or
OFF
(default is
OFF
). The following table lists the default Respiratory limits for
adult, pediatric, and neonate patients in Breaths/Minute, and gives the range in which you can
set these limits:
Using the CO2 Parameter Control Panel
Use the navigation keys to highlight and select the
CO2
numeric display to display the
CO2 Parameter Control Panel:
Figure 9-2 CO2 Parameter Control Menu
On the CO2 Parameter Control Panel, you can select a CO
2
alarm (EtCO2, Breath Rate, or
FiCO2). On the selected alarm settings menu, you can enable/disable alarms and set alarm
limits (as described previously). You can also set the CO
2
sweep speed.
Setting the CO
2
Sweep Speed
The EtCO
2
sweep speed determines the X-axis scale of the capnogram. For patients with
slower respiration rates, a slower sweep speed will make the capnogram easier to view. You can
specify sweep speeds of 3.13, 6.25, and 12.5
mm/second. The default sweep speed is
6.25 mm/second.
Patient Type
Respiration Rate Default
Respiration Rate Range
Adult
Lower: 3 BPM
Upper: 50 BPM
Lower: 0 to 145 BPM
Upper: 5 to 150 BPM
Pediatric
Lower: 3 BPM
Upper: 50 BPM
Lower: 0 to 145 BPM
Upper: 5 to 150 BPM
Neonate
Lower: 12 BPM
Upper: 80 BPM
Lower: 0 to 145 BPM
Upper: 5 to 150 BPM
0 to 30 mmHg
0 to 150 mmHg
CO2 Scale
0 to 100 mmHg
0 to 60 mmHg
CO2
Lower
Upper
EtCO2 Alarm
28
50
No Breath Alarm
FiCO2 Alarm
0
10
30 sec
30
RR/BR Alarm
10
CO2/Resp Sweep Speed
12.5 mm/s
Содержание X Series
Страница 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Страница 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Страница 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Страница 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Страница 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Страница 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Страница 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Страница 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Страница 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Страница 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Страница 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Страница 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Страница 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Страница 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Страница 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Страница 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...